DOI: 10.3724/SP.J.1249.2018.01029

Journal of Shenzhen University Science and Engineering (深圳大学学报理工版) 2018/35:1 PP.29-34

Influence of electrode voltage on imaging performance of streak camera

The newly assembled streak camera can not achieve the optimal imaging at the designed voltage. How to achieve the coordination of multiple electrode voltages is the key to achieve the optimal imaging performance of the streak camera. The relationship between the imaging performance of the streak camera and the electrode voltage is deeply studied based on theoretical analysis and numerical simulation. It is found that the distribution of electrode voltages plays an important role in the imaging performance of the streak camera. The first anode has little effect on the position of the image plane, while the first focusing electrode and the second focusing electrode have significant influence on the image plane, and the adjustment of their voltages has the opposite effect in direction on the image position. The experimental platform was built for testing the streak camera, and the test results are consistent with that of the theoretical analysis. The adjustment of the image plane position and the optimization of the resolution performance can be achieved by the linkage adjustment of the focusing electrode voltage and the anode voltage, which is of great application value for optimizing rapidly the performance for the streak camera.

Key words:electron physics,electron optical devices,streak camera,ultra-fast optics,image tube,laser

ReleaseDate:2018-03-20 15:26:55

[1] 李翔,顾礼,宗方轲,等.高能X射线入射时CsI光阴极的光电发射特性[J].深圳大学学报理工版,2016,33(3):254-258. LI Xiang,GU Li,ZONG Fangke,et al. Photoemission characteristics of a CsI photocathode under high energy X-ray[J]. Journal of Shenzhen University Science and Engineering,2016,33(3):254-258.(in Chinese)

[2] ZONG Fangke, YANG Qinlao, CAI Houzhi, et al. Electron accelerating unit for streak image tubes[J]. Sadhana, 2015, 40(8):2333-2339.

[3] 廖华,杨勤劳.神光Ⅲ原型装置自瞄准X射线扫描相机研制[J].深圳大学学报理工版,2014,31(6):614-617. LIAO Hua,YANG Qinlao.Development of a self targeting X-ray streak camera of Shenguang-Ⅲ prototype facility[J]. Journal of Shenzhen University Science and Engineering,2014,31(6):614-617.(in Chinese)

[4] LI Xiang, GU Li, ZONG Fangke, et al.Temporal resolution limit estimation of x-ray streak cameras using a CsI photocathode[J].Journal of Applied Physics, 2015,118(8):083105.

[5] ZUBER C, BAZZOLI S, BRUNEL P, et al. Picosecond X-ray streak camera dynamic range measurement[J].Review of Scientific Instruments, 2016, 87(9):093501.

[6] 周立伟,公慧,张智诠,等. 两电极静电同心球系统的近轴电子光学及其空间-时间像差[J].物理学报,2010,59(8):5459-5466. ZHOU Liwei, GONG Hui, ZHANG Zhiquan, et al. Paraxial electron optics and its spatial-temporal aberrations for a bi-electrode concentric spherical system with electrostatic focusing[J].Acta Physica Sinica, 2010, 59(8):5459-5466.(in Chinese)

[7] 张敬金,雷保国,杨勤劳.X射线条纹管成像性能提升的研究[J].深圳大学学报理工版,2017,34(1):14-19. ZHANG Jingjin, LEI Baoguo, YANG Qinlao. Improvement of imaging performance for X-ray image tube[J]. Journal of Shenzhen University Science and Engineering, 2017, 34(1):14-19.(in Chinese)